FDA-Device2019-07-31Class II

MAC VU360, Model Number 2030360-001, Electrocardiograph

GE Healthcare, LLC
Hazard

Two issues: 1) The MAC VU360 system may intermittently display an incorrect patient ID or visit number on the screen after scanning the patient s barcode. This may result in the ECG report being assigned to the incorrect patient. 2) The MAC VU360 may have the incorrect patient demographics appear on the patient banner. This may result in the ECG report being assigned to the incorrect patient.

Sold states
Worldwide
Affected count
369 units
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
MAC VU360, Model Number 2030360-001, Electrocardiograph

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2060-2019

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief