FDA-Device2019-08-21Class I

Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.

Medline Industries Inc
Hazard

This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.

Sold states
One distributor located in Illinois.
Affected count
4890
Manufactured in
Three Lakes Drive, N/A, Northfield, IL, United States
Products
Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2067-2019