Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
Excelsior Medical Corp
Hazard
There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.
Sold states
Worldwide Distribution - USA (nationwide) and Canada.
Affected count
4250 tubing sets (PA03 - 78 cases; BX03 - 92 cases (each case contains 25 tubing sets))
Manufactured in
1933 Heck Ave, N/A, Neptune, NJ, United States
Products
Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
Official notice