FDA-Device2021-07-28Class I

Philips Respironics V60 Ventilator with Software Versions 3.00 and 3.10. The Respironics V60/V60 Plus Ventilator is an assist ventilator and is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.

Respironics California, LLC
Hazard

It has been identified that ventilators equipped with High Flow Therapy (Software Versions 3.00 and 3.10), that includes a designed safety mechanism that limits the amount of pressure that can be delivered to a patient, in situations where the system pressure reaches the maximum limit and sounds the "Cannot Reach Target Flow" alarm , the system will reduce pressure and simultaneously decreasing the flow rate to a level below what was set by the clinicians. If the reason for the increase in pressure is not resolved, the ventilator will continue to provide the lower flow rate, which result in patients experiencing oxygen desaturation which can be characterized as moderate or severe hypoxemia.

Sold states
Worldwide Distribution: U.S (nationwide) to states of.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and O.U.S. countries of: Andorra, United Arab Emirates, Albania, Argentina, Austria, Australia, Belgium, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, French Guiana, Guadeloupe, Greece, Hong Kong, Indonesia, Ireland, India, Iceland, Italy, Jordan, Kenya, Kuwait, Lebanon, Latvia, Libya, Morocco, Martinique, Malta, Mexico, Malaysia, New Caledonia, Nigeria, Netherlands, Norway, New Zealand, Philippines, Poland, Portugal, Qatar, R¿union, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, and South Africa.
Affected count
Total of V60/V60 Plus = 23,115 systems
Manufactured in
2271 Cosmos Ct, N/A, Carlsbad, CA, United States
Products
Philips Respironics V60 Ventilator with Software Versions 3.00 and 3.10. The Respironics V60/V60 Plus Ventilator is an assist ventilator and is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2073-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief