FDA-Device2014-07-23Class II

Siemens syngo RT Dosimetrist 2.7 system Product Usage: The intended use of syngo Suite for Oncology Systems is as an accessory to the linear accelerator system to aid and support in the planning and delivery of x-ray radiation for the therapeutic treatment of cancer.

Siemens Medical Solutions USA, Inc
Hazard

An update to the Virtual Simulation software of the syngo RT Dosimetrist is necessary to address a safety issue. In special scenarios within the Structure Operations function the system did not store the selected name from the list of predefined names. After reloading of the data, this structure name had been changed by the system. Additionally, a performance issue was addressed where import of DI

Sold states
US Nationwide Distribution in the states of NH, TN, OH, VA, and TX.
Affected count
7
Manufactured in
51 Valley Stream Pkwy, N/A, Malvern, PA, United States
Products
Siemens syngo RT Dosimetrist 2.7 system Product Usage: The intended use of syngo Suite for Oncology Systems is as an accessory to the linear accelerator system to aid and support in the planning and delivery of x-ray radiation for the therapeutic treatment of cancer.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2075-2014