FDA-Device2012-08-01Class II
Siemens syngo.plaza. Radiological image processing system
Siemens Medical Solutions USA, Inc
Hazard
Siemens became aware of a potential malfunction when using syngo.plaza with software version VA20C_HF01. For datasets with distance measurements it can happen, that all images are not loaded to Viewer but no error message is displayed to user.
Sold states
(USA) nationwide including the states of CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.
Affected count
2
Manufactured in
51 Valley Stream Pkwy, Malvern, PA, United States
Products
Siemens syngo.plaza. Radiological image processing system
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2078-2012Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief