FDA-Device2017-05-24Class II
Lactate Reagent (LACT 2 x 50), Catalog No. A95550 Intended for the quantitative determination of Lactate concentration in human plasma and cerebrospinal fluid (CSF).
Beckman Coulter Inc.
Hazard
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
Sold states
Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.
Affected count
92,837 units total (71,271 units in US)
Manufactured in
250 S Kraemer Blvd, Brea, CA, United States
Products
Lactate Reagent (LACT 2 x 50), Catalog No. A95550 Intended for the quantitative determination of Lactate concentration in human plasma and cerebrospinal fluid (CSF).
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2078-2017Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief