FDA-Device2021-07-28Class II

Hemotherm CE Dual Reservoir Cooler/Heater, Model 400CE, 115V Model, part no. 86022. Used to cool or warm blood during cardiopulmonary bypass procedures

Gentherm Medical, LLC
Hazard

There is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM) to grow in the water systems of any heater-cooler device, and contaminated water from any heater-cooler device has the potential to aerosolize into the operating room during surgery which could lead to patient infection.

Sold states
Worldwide distribution. US nationwide, including Puerto Rico; Argentina, Bangladesh, Chile, Colombia, France, Guatemala, Hungary, India, Indonesia, Israel, Italy, Nepal, Nigeria, Pakistan, Philippines, Romania, Saudi Arabia, Singapore, South Africa, South Korea, Switzerland, Taiwan, Thailand, Turkey, UAE
Affected count
1448 devices (1010 US; 438 OUS)
Manufactured in
12011 Mosteller Rd Fl 3, N/A, Blue Ash, OH, United States
Products
Hemotherm CE Dual Reservoir Cooler/Heater, Model 400CE, 115V Model, part no. 86022. Used to cool or warm blood during cardiopulmonary bypass procedures

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2092-2021