FDA-Device2015-07-29Class II
Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.
Maquet Cardiovascular Us Sales, Llc
Hazard
There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.
Sold states
Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwan and South Africa.
Affected count
12 units
Manufactured in
45 Barbour Pond Drive, N/A, Wayne, NJ, United States
Products
Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2097-2015Don't want to check this manually?