FDA-Device2016-07-06Class II

Dako HER2 CISH pharmDx Kit, product code: SK 109 Product Usage: For in vitro diagnostic use. HER2 CISH pharmDx Kit is intended for dualcolor chromogenic visualization of signals achieved with directly labeled in situ hybridization probes targeting the HER2 gene and centromeric region of chromosome 17. The kit is designed to quantitatively determine HER2 gene status in formalin-fixed, paraffin-embedded breast cancer tissue specimens. Red and blue chromogenic signals are generated on the same

Dako North America Inc.
Hazard

An incorrect HER2 probe concentration was used when producing the probe component (Vial 3) in this kit lot. This has resulted in the HER2 probe concentration being too low.

Sold states
Worldwide Distribution - US Nationwide in the states of: PA, CA, AL, NC and countries of Austria, Bosnia and Herzegovina, Belgium, Bulgaria, Brazil, Canada, Switzerland, Germany, Spain, France, Greece, Italy, Macedonia, Qatar, Romania, Serbia, Sweden, and Turkey.
Affected count
31 kits
Manufactured in
1170 Mark Ave, Carpinteria, CA, United States
Products
Dako HER2 CISH pharmDx Kit, product code: SK 109 Product Usage: For in vitro diagnostic use. HER2 CISH pharmDx Kit is intended for dualcolor chromogenic visualization of signals achieved with directly labeled in situ hybridization probes targeting the HER2 gene and centromeric region of chromosome 17. The kit is designed to quantitatively determine HER2 gene status in formalin-fixed, paraffin-embedded breast cancer tissue specimens. Red and blue chromogenic signals are generated on the same tissue section for evaluation under bright field microscopy. The CISH procedure is automated using Dako Autostainer instruments. HER2 CISH pharmDx Kit is indicated as an aid in the assessment of patients for whom Herceptin (trastuzumab) treatment is being considered. Results from the HER2 CISH pharmDx Kit are intended for use as an adjunct to the clinicopathologic information currently used for estimating prognosis in stage II, node-positive breast cancer patients.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2099-2016

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief