FDA-Device2020-06-03Class II
ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Hazard
If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only be returned to normal operation with the support of a field service engineer
Sold states
US Nationwide distribution.
Affected count
105 systems
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2108-2020More Siemens Medical Solutions USA, Inc recalls
- FDA-Device2026-07-08Siemens ARTIS icono ceiling monitor suspension recalled for loose screw risk
- FDA-Device2026-07-08Siemens ARTIS icono floor recalled for potential movement loss at detector height limits
- FDA-Device2026-07-08Siemens ARTIS icono biplane recalled for potential loss of movement after detector adjustment
- FDA-Device2026-07-08Siemens ARTIS icono ceiling model 11328100 recalled for loss of movement capability
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief