FDA-Device2020-06-03Class II
ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Siemens Medical Solutions USA, Inc
Hazard
If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only be returned to normal operation with the support of a field service engineer
Sold states
US Nationwide distribution.
Affected count
105 systems
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2108-2020Don't want to check this manually?