FDA-Device2020-06-03Class II

VACUETTE Blood Transfer Unit. Lot no. A19024NX. Expiration date: 2022-02-07 - Product Usage: The Blood Transfer Unit is used to transfer blood from a syringe into the evacuated blood collection tube.

C.A. Greiner & Sohne Gesellschaftmbh
Hazard

While transferring blood the needle may dislodged and remain in the tube stopper which may cause needle stick injury or blood leakage.

Sold states
*requested 4/15/2020
Affected count
90,400 pieces.
Manufactured in
Greiner St. #70, N/A, Kremsmunster, N/A, Austria
Products
VACUETTE Blood Transfer Unit. Lot no. A19024NX. Expiration date: 2022-02-07 - Product Usage: The Blood Transfer Unit is used to transfer blood from a syringe into the evacuated blood collection tube.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2116-2020

Don't want to check this manually?