FDA-Device2016-07-06Class II
Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The Ingenuity Core 128 scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes..
Philips Medical Systems (Cleveland) Inc
Hazard
Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
Sold states
US Nationwide Distribution in the states of CA, FL, ID, KS, MI, NY,OH,TX
Affected count
19
Manufactured in
595 Miner Rd, Cleveland, OH, United States
Products
Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The Ingenuity Core 128 scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes..
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2118-2016Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief