FDA-Device2018-06-13Class II
Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 3.2MM X 10MM (002-E-32010) b) D 3.2MM X 12MM (002-E-32012) c) D 3.2MM X 14MM (002-E-32014) d) D 3.2MM X 16MM (002-E-32016) e) D 3.2MM X 18MM (002-E-32018) f) D 3.2MM X 20MM (002-E-32020) g) D 3.2MM X 22MM (002-E-32022) h) D 3.2MM X 24MM (002-E-32024) i) D 3.2MM X 26MM (002-E-32026) j) D 3.2MM X 28MM (002-E-32028) k) D 3.2MM X 30MM (002-E-32030) l) D 3.2MM X 32MM (002-E-32032) m) D 3.2MM X 34MM (
Orthofix, Inc
Hazard
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Sold states
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Affected count
1574
Manufactured in
3451 Plano Pkwy, N/A, Lewisville, TX, United States
Products
Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 3.2MM X 10MM (002-E-32010) b) D 3.2MM X 12MM (002-E-32012) c) D 3.2MM X 14MM (002-E-32014) d) D 3.2MM X 16MM (002-E-32016) e) D 3.2MM X 18MM (002-E-32018) f) D 3.2MM X 20MM (002-E-32020) g) D 3.2MM X 22MM (002-E-32022) h) D 3.2MM X 24MM (002-E-32024) i) D 3.2MM X 26MM (002-E-32026) j) D 3.2MM X 28MM (002-E-32028) k) D 3.2MM X 30MM (002-E-32030) l) D 3.2MM X 32MM (002-E-32032) m) D 3.2MM X 34MM (002-E-32034) n) D 3.2MM X 36MM (002-E-32036) o) D 3.2MM X 38MM (002-E-32038) p) D 3.2MM X 40MM (002-E-32040) q) D 3.2MM X 42MM (002-E-32042) r) D 3.2MM X 44MM (002-E-32044) s) D 3.2MM X 46MM (002-E-32046) t) D 3.2MM X 48MM (002-E-32048) u) D 3.2MM X 50MM (002-E-32050)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2124-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.