FDA-Device2013-09-11Class III

WASP (Walk Away Specimen Process). Automatic Planting and Streaking Instrument Catalog number: W086 WASP is a fully automated system used to process microbiological specimens in liquid phase for the investigation of infectious microbes.

Copan Italia
Hazard

The recall of the WASP (Walk Away Specimen Process) instrument was initiated because Copan Italia SpA has upgraded the software concerning the broth inoculation table feature. The upgrade has been considered necessary from the manufacturer in order to eliminate the very low risk of rotation of the broth table when the door is open.

Sold states
Worldwide Distribution - USA including Oregon and California., and the countries of France, Belgium, Italy, Denmark, Norway, Switzerland, Romania, Finland, Germany, South Africa, Canada, Ireland, UK, Sweden and Japan
Affected count
63 units
Manufactured in
Via Perotti, 10, N/A, Brescia, N/A, Italy
Products
WASP (Walk Away Specimen Process). Automatic Planting and Streaking Instrument Catalog number: W086 WASP is a fully automated system used to process microbiological specimens in liquid phase for the investigation of infectious microbes.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2131-2013