FDA-Device2012-08-08Class II

RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.

Stryker Instruments Div. of Stryker Corporation
Hazard

Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead of 5mm and the cannula is used in a cervical procedure there is a potential for a thermal skin lesion due to a larger lesion creation. If the 10 mm active tip cannula, as opposed

Sold states
Worldwide Distribution - US Nationwide and the countries of England, France, Netherlands, and Switzerland.
Affected count
226
Manufactured in
4100 E Milham Ave, N/A, Portage, MI, United States
Products
RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2135-2012