FDA-Device2020-06-03Class II

Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Outer Diameter 31 mm, Balloon Length 80 mmm, Rx ONLY, CE Marking, Sterile EO, UDI: 00813132024666

Spectranetics Corporation
Hazard

There is a potential for thrombus formation on balloons prepped and staged in the vasculature for long durations. All users, such as those performing lead extraction cases may not be aware of this risk. It was determined that current labeling is insufficient to ensure all potential users minimize the time the balloon remains in vasculature. If left in too long it could lead to embolization.

Sold states
US: Nationwide OUS: Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Japan, Jordan, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, The Netherlands, and the United Kingdom of Great Britain and Northern Ireland
Affected count
N/A
Manufactured in
9965 Federal Dr, N/A, Colorado Springs, CO, United States
Products
Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Outer Diameter 31 mm, Balloon Length 80 mmm, Rx ONLY, CE Marking, Sterile EO, UDI: 00813132024666

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2143-2020