FDA-Device2023-07-19Class II

HeartSine, REF: PAD-PAK-01, PAD-PAK-03 AND PAD-PAK-03j, CE0123 sold separately and with the HeartSine Samaritan Public Access Defibrillator models: 350P/360P/450P/500P

HeartSine Technologies Ltd
Hazard

Single use battery and electrode cartridge may be rendered inoperable due to depleted battery cells which could prevent the defibrillator from analyzing patient condition or delivering therapy correctly.

Sold states
Worldwide Distribution: US (nationwide): AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA & WI and OUS (countries of): Australia, Canada, Denmark, France, Iceland, Ireland, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Peru, Poland, Singapore, South Africa, Spain, Thailand, & United Kingdom.
Affected count
30,022 units
Manufactured in
207 Airport Road West, N/A, Belfast, N/A, United Kingdom
Products
HeartSine, REF: PAD-PAK-01, PAD-PAK-03 AND PAD-PAK-03j, CE0123 sold separately and with the HeartSine Samaritan Public Access Defibrillator models: 350P/360P/450P/500P

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2143-2023

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief