FDA-Device2024-06-26Class II
Centrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient support system to prevent and/or treat pressure injuries.Multiple model numbers of Centrella Max mattresses and pro+ mattresses. (1) Centrella Max Surface, REF P7922A01; (2) Centrella Max Surface, REF P7922A02; (3) Centrella Max Surface with X-Ray, REF P7922A03; (4) Centrella Max,Narrow, Rental, REF P7922ARENT01; (5) Centrella Pro+ 36" Surface with X-Ray, REF P7923A01; (6) Centrella Pro+
Baxter Healthcare Corporation
Hazard
There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.
Sold states
Distribution was made nationwide. There was also government/military distribution. Foreign distribution was made to Australia, Canada, China, Japan, Qatar, Saudi Arabia, Singapore, Taiwan, Turkey, and United Arab Emirates.
Affected count
38,136 mattress units and 20,187 service part units
Manufactured in
1 Baxter Pkwy, N/A, Deerfield, IL, United States
Products
Centrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient support system to prevent and/or treat pressure injuries.Multiple model numbers of Centrella Max mattresses and pro+ mattresses. (1) Centrella Max Surface, REF P7922A01; (2) Centrella Max Surface, REF P7922A02; (3) Centrella Max Surface with X-Ray, REF P7922A03; (4) Centrella Max,Narrow, Rental, REF P7922ARENT01; (5) Centrella Pro+ 36" Surface with X-Ray, REF P7923A01; (6) Centrella Pro+ 40" Surface with X-Ray, REF P7923A02; (7) Centrella Pro+36" Surface, REF P7923A03; (8) Centrella Pro+ 40" Surface, REF P7923A04; (9) Pro+ MRS Surface for Hillrom 900 with X-Ray, REF P006800A01; (10) 80" MRS with X-Ray AU/NZ REF P006800A03; (11) Pro+ MRS Surface for Versacare with X-Ray, REF P3255A01; (12) Pro+ MRS Surface for Versacare, REF P3255A02; (13) Pro + MRS Surface for Versacare with X-Ray, REF P3255ARENT01; (14) Pro+ 36"MRS Surface with X-Ray, REF P7924A01; (15) Pro+ 40" MRS Surface with X-Ray, REF P7924A02; (16) Pro+ 36" MRS Surface, REF P7924A03; (17) Pro+ 40" MRS Surface, REF P7924A04; (18) Pro+ 36" MRS Rental Surface with X-Ray, REF P7924ARENT01.