FDA-Device2025-07-30Class II

Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP, 10005941JPN

Covidien
Hazard

Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting

Sold states
Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR, WA, SD, KY, DE, NJ, UT, NV, MA, MT, GU, NE, LA, HI, OR, RI, NH, ND, NM, ME, CT, DC, WY and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bhutan, Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic Of, Ireland, Isle of Man, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, Zimbabwe.
Affected count
264,790
Manufactured in
6135 Gunbarrel Ave, N/A, Boulder, CO, United States
Products
Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP, 10005941JPN

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2151-2025