Pausch Uroview FD fluoroscopic X-ray system UDI number (01)04260491620040 Primarily for urological applications (functional X-ray diagnostics, endourology and minimal invasive urology / surgery). The system, which includes a radiologic / urologic treatment table, maybe used for urological treatment, planning and diagnostic procedures, including but not limited to.
Hazard
During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if footswitch + exposure + first primary barrier are closed the signal will be transmitted to the generator for activating X-ray. During intended use only with this footswitch X-ray should be activated.
Sold states
Nationwide Distribution
Affected count
10 systems
Manufactured in
Graf Zeppelin Strasse 1, Erlangen, Germany
Products
Pausch Uroview FD fluoroscopic X-ray system UDI number (01)04260491620040 Primarily for urological applications (functional X-ray diagnostics, endourology and minimal invasive urology / surgery). The system, which includes a radiologic / urologic treatment table, maybe used for urological treatment, planning and diagnostic procedures, including but not limited to.