FDA-Device2021-08-04Class II

Proteus 235-Proton Therapy System

Ion Beam Applications S.A.
Hazard

if the user forgets to select the trigger input manually while beam gating is prescribed for the patient, the proton irradiation will be delivered without gating. If the beam is not gated as prescribed, the treatment field may be delivered with an error in position

Sold states
FL, PA, TN, TX, VA Foreign: Germany, Italy, Russia, Sweden , India, Netherlands
Affected count
5 US; 8 OUS
Manufactured in
Chemin Du Cyclotron 3, N/A, Louvain-la-neuve, N/A, Belgium
Products
Proteus 235-Proton Therapy System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2155-2021