FDA-Device2024-06-26Class II
NxStage PureFlow-B Solution, Premixed Dialysate for Hemodialysis, Device Catalog Numbers RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-405, RFP-406, RFP-453, RFP-454, RFP-456
NxStage Medical Inc
Hazard
The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.
Sold states
US Nationwide distribution.
Affected count
1316936 units
Manufactured in
350 Merrimack St, N/A, Lawrence, MA, United States
Products
NxStage PureFlow-B Solution, Premixed Dialysate for Hemodialysis, Device Catalog Numbers RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-405, RFP-406, RFP-453, RFP-454, RFP-456
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2160-2024Don't want to check this manually?