FDA-Device2016-07-20Class II
Target Detachable Coils: TARGET HELICAL NANO 2MM X 3CM MODEL Number:M0035452030 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Stryker Neurovascular
Hazard
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
Sold states
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
Affected count
45,302 (8,188 in the U.S.) - total, all model numbers
Manufactured in
47900 Bayside Pkwy, Fremont, CA, United States
Products
Target Detachable Coils: TARGET HELICAL NANO 2MM X 3CM MODEL Number:M0035452030 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2179-2016Don't want to check this manually?