FDA-Device2014-08-20Class II

VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS¿ CA 19-9 Reagent Packs are used on VITROS¿ ECi/ECiQ Immunodiagnostic Systems, VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600 Integrated Systems for the quantitative measurement of 1116-NS-19-9 defined antigen in human serum and plasma.

ORTHO-CLINICAL DIAGNOSTICS
Hazard

Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS¿ CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.

Sold states
Worldwide Distribution - USA (nationwide) and Internationally to France and Japan
Affected count
475
Manufactured in
FELINDRE MEADOWS, N/A, Bridgend, N/A, United Kingdom
Products
VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS¿ CA 19-9 Reagent Packs are used on VITROS¿ ECi/ECiQ Immunodiagnostic Systems, VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600 Integrated Systems for the quantitative measurement of 1116-NS-19-9 defined antigen in human serum and plasma.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2183-2014