FDA-Device2021-08-11Class II

REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569

Abbott Laboratories
Hazard

There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results.

Sold states
Worldwide distribution - US Nationwide distribution in the states of VA, UT, TX, TN, SD,PA, NY, ME, ID, CA and the countries of AFGHANISTAN, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVI, CANADA, CAYMAN ISLANDS, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FAROE ISLAND, FINLAND, FRANCE, GERMANY, GREENLAND, Hong Kong, INDIA, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, MONTENEGRO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, PORTUGAL, PR CHINA, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UNITED KINGDOM, VIETNAM.
Affected count
4,727 units
Manufactured in
1915 Hurd Dr, Irving, TX, United States
Products
REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2184-2021