FDA-Device2016-07-27Class II

Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System

Toshiba American Medical Systems Inc
Hazard

During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the same. It incorrectly used in the calculation a dose value from a previous exposure.

Sold states
US Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE.
Affected count
23
Manufactured in
PO Box 2068, 2441 Michelle Dr, Tustin, CA, United States
Products
Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2188-2016