GE Healthcare Advantage Workstations, Advantage Worksation version 4.2 and 4.3 with Volume Viewer 2 (version 6.4-6.8) provided with Advance Vessel Analysis (AVA) and Advantage Workstation 4.4 with Volume Viewer 2 (version 6.8 or before) or Volume Viewer 3 (version 7.2 -7.3) provided with Advanced Vessel Analysis (AVA), Model 2378698-2 Product Usage: Advantage Workstation version 4.2: The Advantage Workstation 4.2 is a review station, which allows easy selection, review, processing, filming a

GE Healthcare, LLC
Hazard

GE Healthcare is informing you of incorrect measurements when using the Aorta analysis protocol or customized protocols. Until a new software version is installed they are asking you not to use the aorta analysis protocol or the customized protocols for an aneurysm. A second issue was also discovered internally. When loading PET data into the Volume Viewer, if slices are missing in a PET serie

Sold states
Worldwide Distribution - US (nationwide) including the states of : AL, AZ, AR, CA, CO, DE, DC, FL, GA, GU, HI, IK, IL, IN, IA, KY, LA, MD, MA, MI, MN, MS, MO,MT, NE,NH,NJ, NM, NY, NC, ND, OH, OK, PA, PR, SC, SD, TN, TX, VA, WA, WV, WI and the countries of YUGOSLAVIA, VIETNAM, VENEZUELA, UNITED KINGDOM, UNITED ARAB EMIRATES, UKRAINE, TURKEY, TUNISIA, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SINGAPORE, SA¿DE ARABIA, RUSSIAN FEDERATION, ROMANIA, REPUBLIC OF MALTA, PORTUGAL, POLAND, PHILIPPINES, NORWAY, NIGERIA, NEW ZEALAND, MEXICO, KOREA, KENYA, JAPAN, ITALY, IRELAND, INDONESIA, INDIA, HUNGRY, HONG KONG, GREECE, GERMANY, FRANCE, FINLAND, EGYPT, DENMARK, CZECH REPUBLIC, CHINA, CHILE, CANADA, BULGARIA, BRAZIL, BELGIUM, BELARUS, AUSTRIA, AUSTRALIA, ARGENTINA, ALGERIA.
Affected count
1055
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
GE Healthcare Advantage Workstations, Advantage Worksation version 4.2 and 4.3 with Volume Viewer 2 (version 6.4-6.8) provided with Advance Vessel Analysis (AVA) and Advantage Workstation 4.4 with Volume Viewer 2 (version 6.8 or before) or Volume Viewer 3 (version 7.2 -7.3) provided with Advanced Vessel Analysis (AVA), Model 2378698-2 Product Usage: Advantage Workstation version 4.2: The Advantage Workstation 4.2 is a review station, which allows easy selection, review, processing, filming and medial interchange of multi-modality images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed images may be used as an element for diagnosis. Advantage Workstation version 4.3/4.4 : The Advantage Workstation 4.3 is a review station, which allows easy selection, review, processing, filming and medial interchange of multi-modality DIACOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed images of the AW monitor may be used as a basis for diagnosis, except in the case of mammography.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2203-2012