FDA-Device2014-08-20Class II

Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to conduct syncope examinations.

Medical Positioning Inc
Hazard

Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.

Sold states
Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.
Affected count
17 beds
Manufactured in
1717 Washington St, N/A, Kansas City, MO, United States
Products
Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to conduct syncope examinations.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2210-2014