FDA-Device2012-08-22Class II
Trabecular Metal Reverse Shoulder Liner 40mm, part 00-4309-028-00 - Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."
Zimmer, Inc.
Hazard
Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
Sold states
Worldwide Distribution - USA (nationwide) and internationally to Germany, Belgium, Denmark, Finland, France, Italy, Netherlands, Spain, Sweden, Switzerland, and the United Kingdom.
Affected count
185 (162 in USA)
Manufactured in
345 E Main St, Warsaw, IN, United States
Products
Trabecular Metal Reverse Shoulder Liner 40mm, part 00-4309-028-00 - Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2212-2012Don't want to check this manually?