FDA-Device2013-09-25Class II
Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems: The Neptune Waste management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautory or laser devices. ***The single-port manifold for the Stryker Neptune 1 (part # 700-015-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bron
Zeppessis Reprocessing, LLC
Hazard
Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems are recalled due to lacking 510K and and without a validated cleaning protocol for reprocessing the used manifolds.
Sold states
Distributed in the states of California, Idaho, Oregon, and Washington.
Affected count
undetermined the distributed amounts
Manufactured in
121 E 38th St Ste 101, N/A, Garden City, ID, United States
Products
Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems: The Neptune Waste management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautory or laser devices. ***The single-port manifold for the Stryker Neptune 1 (part # 700-015-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 1 (part # 700-020-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***single-port manifold for the Stryker Neptune 2 (part # 702-025-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 2 (part # 702-020-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2212-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief