FDA-Device2023-08-02Class II
REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. D684-00-0403-05 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. D684-00-0403-06 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. D684-00-0403-10
Datascope Corp.
Hazard
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Sold states
Domestic distribution nationwide. International distribution worldwide.
Affected count
321,609 total kits
Manufactured in
15 Law Dr Unit 1, N/A, Fairfield, NJ, United States
Products
REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. D684-00-0403-05 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. D684-00-0403-06 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. D684-00-0403-10
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2224-2023Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.