FDA-Device2017-06-14Class II
Sesamoid Plasty, Stereotaxic Instrument The Sesamoid Plasty workstation is used to run the tracking software as part of the software based navigation systems. The affected Sesamoid Plasty components are the central structural column and its base that hold up the computer and the tracking camera on the Sesamoid Plasty workstation. They are also designed to be disassembled such that the column can be folded onto itself for ease of transport after use. This is a retrospective recall report for a r
Orthosoft, Inc. dba Zimmer CAS
Hazard
Zimmer CAS voluntarily conducted a retrospective recall of the Sesamoid Plasty due to the risk of the top of the workstation to potentially fall over during disassembly.
Sold states
Worldwide Distribution - US including FL, CA Foreign: Austrailia, New Zealand, France, Brazil
Affected count
10
Manufactured in
75 Queen St #3300, N/A, Montreal, N/A, Canada
Products
Sesamoid Plasty, Stereotaxic Instrument The Sesamoid Plasty workstation is used to run the tracking software as part of the software based navigation systems. The affected Sesamoid Plasty components are the central structural column and its base that hold up the computer and the tracking camera on the Sesamoid Plasty workstation. They are also designed to be disassembled such that the column can be folded onto itself for ease of transport after use. This is a retrospective recall report for a removal initiated on 17 June 2008
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2227-2017Don't want to check this manually?