FDA-Device2014-08-20Class II
23.5 Deg Rasp Handle Assy- RIGHT Nonsterile
Zimmer, Inc.
Hazard
The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring.
Sold states
Worldwide Distribution-USA (nationwide) including the states of AK, AL, AZ, CA, CO, FL, GA, IN, KS, KY, MI, MN, NC, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, and VA, and the countries of Australia, France, Germany, Italy, Switzerland, and Canada.
Affected count
N/A
Manufactured in
1800 W Center St, N/A, Warsaw, IN, United States
Products
23.5 Deg Rasp Handle Assy- RIGHT Nonsterile
Official notice