FDA-Device2016-08-03Class II
GE Inspection Technologies, LP - Phoenix v|tome|x m240 cabinet x-ray systems. A microfocus CT system used primarily for 3D metrology and analysis.
GE Inspection Technologies, LP
Hazard
Radiation emissions which exceeded the limit in the performance standard were detected during maintenance of one system. The identified cause was insufficient attachment of shielding in the door. Jostling during delivery was identified as the probable cause of this failure to comply with the performance standard.
Sold states
US Distribution to the states of : FL, TN, and TX.
Affected count
4
Manufactured in
50 Industrial Park Rd, N/A, Lewistown, PA, United States
Products
GE Inspection Technologies, LP - Phoenix v|tome|x m240 cabinet x-ray systems. A microfocus CT system used primarily for 3D metrology and analysis.