FDA-Device2024-07-10Class II

Acumatch L-Series BIPOLAR LINER, 22mm (BIPOLAR ENDOPROSTHESIS-Hip), Item Numbers: a) 100-22-19, SZ. L; b) 100-22-20, SZ. M; c) 100-22-21, SZ. N; d) 100-22-22, SZ. P; e) 100-22-23, SZ. R; f) 100-22-24, SZ. S; g) 100-22-25, SZ. T

Exactech, Inc.
Hazard

The AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), Nylon, and EVOH, or 2) LDPE and Nylon without EVOH.

Sold states
Worldwide - US distribution in Puerto Rico and the countries of Great Britain, Japan.
Affected count
589 units
Manufactured in
2320 Nw 66th Ct, N/A, Gainesville, FL, United States
Products
Acumatch L-Series BIPOLAR LINER, 22mm (BIPOLAR ENDOPROSTHESIS-Hip), Item Numbers: a) 100-22-19, SZ. L; b) 100-22-20, SZ. M; c) 100-22-21, SZ. N; d) 100-22-22, SZ. P; e) 100-22-23, SZ. R; f) 100-22-24, SZ. S; g) 100-22-25, SZ. T

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2236-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief