Siemens Dimension Chemistry Wash (RD701) is an accessory of the flex cartridge system used on the Siemens Dimension RxL and Xpand assays. The wash buffer is used to clean the probe sensor between uses and prepares it for taking the next reading/measurement.

Siemens Healthcare Diagnostics, Inc.
Hazard

Siemens Healthcare Diagnostics has received complaints for the Dimension Chemistry Wash (RD701) of QC and patient sample result accuracy shifts on Troponin I, CTNI (RF421C), LTNI (RF521C); Free T4, FT4 (RF410); Thyroid Stimulating Hormone, TSH (RF412); and N-terminal pro-brain natriuretic peptide, PBNP (RF423A) and LPBN (RF523A) when using Dimension® Chemistry Wash lot RD41041. Internal inv

Sold states
USA (nationwide) Distribution and Internationally to Canada.
Affected count
3200
Manufactured in
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, United States
Products
Siemens Dimension Chemistry Wash (RD701) is an accessory of the flex cartridge system used on the Siemens Dimension RxL and Xpand assays. The wash buffer is used to clean the probe sensor between uses and prepares it for taking the next reading/measurement.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2238-2014