FDA-Device2018-06-27Class II
G-PIN 2.8BL 3.2PL 98CM 2PK; 467260 NL DRIVING GUIDE 4.4MM X 80CM; 467220 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
Zimmer Biomet, Inc.
Hazard
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.
Sold states
Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.
Affected count
11
Manufactured in
56 E Bell Dr, N/A, Warsaw, IN, United States
Products
G-PIN 2.8BL 3.2PL 98CM 2PK; 467260 NL DRIVING GUIDE 4.4MM X 80CM; 467220 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2238-2018Don't want to check this manually?