FDA-Device2012-08-29Class II
GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused protection of internal system components from the high voltage system components, such as the X-ray tube or High Voltage Transformer. System model number A349855, Snubber Board number 00-880405-03 rev 5. This product is used for image intensified fluoroscopic x-rays for interventional radiologic procedures.
GE OEC Medical Systems, Inc
Hazard
GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
Sold states
Worldwide and Nationwide (USA) distribution.
Affected count
845 units
Manufactured in
384 Wright Brothers Dr, N/A, Salt Lake City, UT, United States
Products
GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused protection of internal system components from the high voltage system components, such as the X-ray tube or High Voltage Transformer. System model number A349855, Snubber Board number 00-880405-03 rev 5. This product is used for image intensified fluoroscopic x-rays for interventional radiologic procedures.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2241-2012Don't want to check this manually?