FDA-Device2024-07-10Class II

WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc. REF Numbers: T0449-01

WOM World Of Medicine AG
Hazard

Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery

Sold states
US Nationwide distribution.
Affected count
1700 units
Manufactured in
Alte Poststr. 11, N/A, Ludwigsstadt, N/A, Germany
Products
WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc. REF Numbers: T0449-01