FDA-Device2021-08-18Class II

FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127

BioFire Diagnostics, LLC
Hazard

Due to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error.

Sold states
U.S. Nationwide Distribution in the states of : FL, MI, NJ, OH, VA and WI. O.U.S.: None
Affected count
29 kits
Manufactured in
515 S Colorow Dr, N/A, Salt Lake City, UT, United States
Products
FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2245-2021