FDA-Device2020-06-10Class II

ProxiDiagnost N90, UDI 706100

Philips North America, LLC
Hazard

The thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first failure and the thermo switch is not activated, the device may begin to generate smoke.

Sold states
Domestic distribution nationwide. Foreign distribution to Chile and Germany.
Affected count
91
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
ProxiDiagnost N90, UDI 706100

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2256-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief