FDA-Device2014-08-27Class II

Tibila Insert CK, Balanced Knee System Revision, REF563-1YYY Product Usage: UHMWPE component used in TKA and revision TKA procedures to serve as a tibial bearing surface for articulation with the femoral component where additional joint constraint is needed

Ortho Development Corporation
Hazard

Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.

Sold states
Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
Affected count
1
Manufactured in
12187 Business Park Dr, Draper, UT, United States
Products
Tibila Insert CK, Balanced Knee System Revision, REF563-1YYY Product Usage: UHMWPE component used in TKA and revision TKA procedures to serve as a tibial bearing surface for articulation with the femoral component where additional joint constraint is needed

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2257-2014

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief