FDA-Device2025-08-13Class II
DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
Hazard
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
Sold states
Worldwide distribution - US Nationwide.
Affected count
1 system
Manufactured in
1000 Lake Hazeltine Dr, N/A, Chaska, MN, United States
Products
DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2267-2025More Beckman Coulter, Inc. recalls
- FDA-Device2026-04-15Beckman Coulter MicroScan Neg MIC 3J antibiotic susceptibility panels recalled for manufacturing defects
- FDA-Device2026-04-15MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
- FDA-Device2026-04-01Beckman Coulter MicroScan Neg MIC 56 panels recalled for potential well contamination
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief