FDA-Device2025-08-13Class II

DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461

Beckman Coulter, Inc.
Hazard

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

Sold states
Worldwide distribution - US Nationwide.
Affected count
1 system
Manufactured in
1000 Lake Hazeltine Dr, N/A, Chaska, MN, United States
Products
DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2270-2025

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief