FDA-Device2021-08-18Class II
Arcos Modular Revision Hip System High Offset Cone Prox Body, Size C, 70 MM Item Number: 11-301333
Biomet, Inc.
Hazard
Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention
Sold states
US Nationwide distribution.
Affected count
12 units
Manufactured in
56 E Bell Dr, N/A, Warsaw, IN, United States
Products
Arcos Modular Revision Hip System High Offset Cone Prox Body, Size C, 70 MM Item Number: 11-301333
Official notice