FDA-Device2013-10-02Class II
BD Epilor Syringe BD Franklin Lakes, NJ 07417 BD EDC Laagstraat 57, B-9140 Temse-Belgium The Epilor syringe is used in conjunction with an epidural needle with the purpose of finding the epidural space.
Becton Dickinson & Company
Hazard
BD has received several reports indicating that the plunger of the 7 mL BD Epilor plastic BD Luer-Lok Loss of Resistance Syringe may stall or stick within the syringe preventing the user from detecting a loss of resistance upon entering the epidural space.
Sold states
Worldwide Distribution - USA including FL, VA, TX, NY, IL, AZ, MI, RI, NJ, SC, MO, TN, CA, MN, UT, GA, NE, OH, MD, PA, MT and internationally to Brazil, Australia, Canada, and Puerto Rico..
Affected count
1,335,885 units
Manufactured in
1 Becton Dr, N/A, Franklin Lakes, NJ, United States
Products
BD Epilor Syringe BD Franklin Lakes, NJ 07417 BD EDC Laagstraat 57, B-9140 Temse-Belgium The Epilor syringe is used in conjunction with an epidural needle with the purpose of finding the epidural space.
Official notice