FDA-Device2016-08-03Class II
Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile Catalog Numbers 2199-2022, 2199-2028 and 2199-2032. For use in total hip arthroplasty.
Stryker Howmedica Osteonics Corp.
Hazard
Impactors do not meet the required sterility assurance level (SAL) of 10-6 when using the recommended steam sterilization method.
Sold states
Worldwide distribution. US nationwide (AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OH, PA, TX, WA, WI, WV), Sweden, South Africa, Chile, Australia, New Zealand, Canada, United Kingdom, Switzerland, Colombia, and Poland.
Affected count
1,864 units
Manufactured in
325 Corporate Dr, N/A, Mahwah, NJ, United States
Products
Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile Catalog Numbers 2199-2022, 2199-2028 and 2199-2032. For use in total hip arthroplasty.
Official notice