FDA-Device2014-08-27Class II
PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilization case is to organize, store and protect its contents from damage consistent with typical hospital or surgical center circulation (i.e. between central services and the operating room) and during transport between a Zimmer distributorship and a health care facility by Zimmer distributorship personnel.
Zimmer, Inc.
Hazard
The firm discovered that the current Personal Tibial/General Fixed Bracket Trays may not have the specific geometry to accommodate the tibial drill and stop guide with proper position/orientation. The drill bracket and drill guide are not the correct geometry to allow for a proper fit in the Persona Fixed Case Kit which could lead to improper sterilization of the tibial drill and stop guide.
Sold states
Worldwide Distribution - USA including AZ, CA, CO, FL, IL, IN, KS, MI, MN, NC, NJ, NY, OH, OR, PA, TN, TX & VA and Internationally to AUSTRALIA, FRANCE, INDIA, ITALY, and GERMANY.
Affected count
190 units
Manufactured in
1800 W Center St, Warsaw, IN, United States
Products
PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilization case is to organize, store and protect its contents from damage consistent with typical hospital or surgical center circulation (i.e. between central services and the operating room) and during transport between a Zimmer distributorship and a health care facility by Zimmer distributorship personnel.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2288-2014Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief