FDA-Device2019-08-21Class II
ARIES Extraction Kit; ARIES System; SYNCT , REF 50-10026, UDI # 00840487100073
Luminex Corporation
Hazard
Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
Sold states
US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
Affected count
N/A
Manufactured in
12212 Technology Blvd, N/A, Austin, TX, United States
Products
ARIES Extraction Kit; ARIES System; SYNCT , REF 50-10026, UDI # 00840487100073